We present measurements of the spectral response of the four Advanced ACTPol (AdvACT) multichroic detector arrays, which observed in frequency bands centered near 30, 40, 90, 150, and 220 GHz. The passbands were measured with a Fourier transform spectrometer (FTS) combined with coupling optics that direct and match the output of the FTS to the AdvACT receiver. We perform optical simulations of the FTS and coupling optics in order to model frequency-dependent systematic effects in this system, examining the systematics of each individual measurement component in detail. We use these simulations to apply corrections to the measurements and evaluate the remaining systematic uncertainty. We combine these systematic estimates with statistical errors to determine the measured passbands and their uncertainties from our FTS measurement, which are typically $0.75$ to $1.5%$ per detector array for the central frequency. We present the passband characteristics for all measurements made by ACT in a common framework. Additionally, we discuss in detail limits set by the systematics in our measurements as well as ways to minimize the systematics of each component to improve passband measurements in the future.