Laboratory Characterization of a Ge-CCD Prototype for Future Near-Infrared Astronomical Instrumentation

Kavli Affiliate: Sven Herrmann
| Summary:
Germanium charge-coupled devices (Ge-CCDs) extend CCD imaging beyond the silicon cutoff and are candidate detectors for the near-infrared channel of future astronomical instruments, in particular the spectrographs of next-generation wide-field surveys. We report the laboratory characterization at Brookhaven National Laboratory (BNL) of a prototype fabricated at MIT Lincoln Laboratory. We describe the cryogenic test system and readout, and present measurements of the output waveform, dark current, readout noise, and X-ray response. The dark current at 94.69 K is $26.12 pm 0.08$ e$^-$ pixel$^-1$ s$^-1$. Under $^55$Fe irradiation without on-chip binning, the charge generated by an X-ray is confined within $2times2$ pixels, and the Mn K$α$ line has a full width at half maximum (FWHM) of $8.5 pm 0.1$% for single-pixel events. Since the charge generated by X-rays above about 30 keV spreads over $3times3$ or more pixels, the CCD was operated with $2times2$ on-chip binning at higher energies. This yielded FWHMs of $8.4pm0.4$% at 30.9 keV ($^133$Ba, ASCA grades 0, 2, 3, 4, and 6) and $7.45pm0.04$% at 59.5 keV ($^241$Am, all grades). These results establish a baseline for the device and for the BNL test system. Measurements of the temperature dependence, optical response, and charge-transfer performance will connect this baseline to the requirements of near-infrared astronomical instruments.
| Search Query: arXiv Query: search_query=au:”Herrmann Sven”&id_list=&start=0&max_results=10